Altana Byko-Test MP0R Manuel d'utilisateur Page 36

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 42
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 35
36
Normalization performed by mistake on calibration
standard instead of on uncoated specimen.
Perform normalization on uncoated specimen.
Er14 Internal instrument error: Not able to compute
master characteristic. The original master
characteristic is retained.
Repeat master calibration.
If this error occurs repeatedly:
Contact customer service.
Er15 Not able to save the master characteristic.
Probe may be defective.
Contact customer service.
Er17 Not enough readings. Thus, not able to finish a
block (applies to measurements w. fixed block
limits).
Make additional measurements until the first
block can be finished. Then generate a new final
result.
Er22 Internal instrument error. Error accessing the
measurement memory or. EEPROM.
Contact customer service.
Vue de la page 35
1 2 ... 31 32 33 34 35 36 37 38 39 40 41 42

Commentaires sur ces manuels

Pas de commentaire