Altana Byko-Test MP0R Manuel d'utilisateur Page 17

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 42
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 16
17
8 Calibration
The following items are required for performing a calibration: An uncoated specimen: In geometry & substrate
material, this specimen should correspond exactly to the parts to be measured and a calibration standard
(instrument-specific correction foil 75 µm (3 mils)).
Note: All readings in the memory will be deleted when performing a calibration.
Calibrating the instrument (Prerequisite: Instrument must be turned ON)
1. Select the menu
with button [] or [].
2. Press button [OK]. The message "Base" (for “substrate (base) material”) appears.
3. Perform about 5 measurements on the uncoated specimen. After each measurement, the current reading will
appear on the display. (continued on next page!)
4. Press button [OK].
By pressing button [OK], the mean value of all readings will be calculated and used for normalization.
Vue de la page 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 41 42

Commentaires sur ces manuels

Pas de commentaire