color-view™ Spectrophotometer User’s Guide
56
3 Specimen plane and aperture can be cleaned with any
common industrial cleaning solvent (avoid contact of
painted surface and interior of the sensor with solvent).
Use of pressurized air is to be avoided as the inner
components of the sensor can be damaged.
3.4.4 Fuse Replacement Procedures
To replace the fuse:
1 Turn the Power switch on the unit off and remove power
cord.
2 With a fingernail or a small flat-blade screwdriver, gently
open the fuse box by inserting it in the small groove next
to fuse compartment door (Figure 3-3) and exerting
lateral pressure.
3 Take out fuses and replace with replacement fuses.
WARNING
Note the following label located near the fuse box:
CAUTION: FOR CONTINUED PORTECTION AGAINST RISK
OF FIRE, REPLACE ONLY WITH SAME TYPE AND RATING
OF FUSE SHOWN.