temp-gard basic Hardware
6
Specifications of the temp-gard basic System with printer (canon)
Channels 8
Thermocouple Type K
Temperature Range -20° to +500 °C (*1)
Sampling Interval 1 second to 60 minutes
Accuracy ± 0,5 °C
Resolution 0,1 °C
Manual Trigger Yes
Time Trigger Yes (0 – 3000 min.)
Temperature Trigger Yes (+30° to +100 °C)
Battery 2 x 1.5 V Alkaline (LR6 AA)
Battery Life (Measuring Time) > 200 hours at 5 sec interval
Length 138 mm
Width 147 mm
Height 38 mm
Weight 0,7 kg
(*1) The practical maximum temperature is limited by the temperature capabilities of the
thermal barrier. If the temp-gard basic system is used with the standard thermal
barrier, a maximum temperature of +300 °C must not be exceeded.
Data Storage
Time > 2 years on PCMCIA card
Storage
Examples:
measurement with 4 sensors
measurement with 4 sensors
measurement with 8 sensors
measurement with 8 sensors
512 kB on PCMCIA card
Examples:
1 s interval = 12.8 h
5 s interval = 64 h
1 s interval = 7.2 h
5 s interval = 36 h