11.1.1.2 SMC - Statistical Measurement Control
Highlight the function with the scroll wheel. Pressing
the wheel activates or deactivates the function. If
SMC is activated a checkmark appears.
For SMC mode you must preset a minimum number
of measurements ( n > 3 ). To keep the required
number of measurements as low as possible, a
default standard error is stored in the instrument. In
case of uniform surfaces, the standard error is often
reached after the minimum number of
measurements. In case of non-uniform surfaces,
e.g. problem areas, additional measurements will be
required by the instrument.
11.1.2 Color
This function lets you choose the measurement
angles, scales and illuminant/observer.
Setup