Altana LC 2 Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Instruments de mesure Altana LC 2. ALTANA LC 2 User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer

Résumé du contenu

Page 1 - Leitfähigkeitsmessgerät LC 2

Additives & InstrumentsA member of Operating InstructionsBetriebsanleitungIstruzioni d’usoMeasure what you see.Conductivity Meter LC 2Leitfähigke

Page 2

Please note An automatic self-test is carried out with every measurement. During the measurement all segments are displayed and the battery voltage i

Page 3

4. Evaluation The measured value constitutes the resistance R [M] of the liquid between the electrodes (in the annular gap). The cell constant whic

Page 4 - Conductivity Meter

5. Calibration The instrument is preset to a factor of 1.000 (i.e.1 M or 1.00 S). For calibration, a resistor of 1 M ± 1 % or a probe and a cal

Page 5

If a button is pressed constantly, the value is automatically reduced or increased. A measurement range between 0.900 and 1.100 is possible. It corre

Page 6

6. Trouble Shooting Symptom On starting the measurement (pressing the measuring button) the display shows "Io.bA". Possible Cause and Rem

Page 7 - 1. Safety Instructions

Symptom After completion of the 10 sec. preparation time, the displaycontinues to show "overflow". Possible Cause and Remedy • Check conne

Page 8 - 2. System Description

7. Ordering Guide / Technical Specification Cat.No. PW-1710 Conductivity Tub Electrode with connection cable ASTM D 5682

Page 9

Cat.No. PW-1722 Conductivity Meter BYK LC 2 with digital display ASTM D 5682 Measuring 0.05 -19.99 M Range: 19.99 - 0.05 S Resolution: 0.01

Page 10

198 013 256 EDI 0806 Leitfähigkeitsmessgerät LC 2 Best.-Nr. PW-

Page 12 - 5. Calibration

198 013 256 EDI 0806 Conductivity Meter LC 2 Cat. No. PW-1722,

Page 13

Leitfähigkeitsmessgerät LC 2 Best.Nr. PW-1722, PW-1710, PW-1712 Inhaltsverzeichnis • Sicherheitshinweise • Systembeschreibung •

Page 15

Leitfähigkeits-Messzelle Bestell-Nr. PW-1710 Leitfähigkeits-Messgerät LC 2 Bestell-Nr. PW-1722 Leit

Page 16

1.Sicherheitshinweise • Das Instrument darf nicht in ex-geschützten Betriebsstätten verwendet werden! • Verwenden Sie

Page 17 - 55 x 105 x 145 mm

2. Systembeschreibung Das Leitfähigkeitsmessgerät LC 2, ASTM D 5682, dient zur Prüfung von Flüssigkeiten mit geringer

Page 18

3. Inbetriebnahme/ Durchführung einer Messung • Messzelle auf Sauberkeit prüfen und zusammen-stecken. Durch die massive Ausführu

Page 19

Bitte beachten Bei jeder Messung wird automatisch ein Selbsttest durchgeführt. Hierbei werden für die Dauer der Messung alle Segmente eingeschaltet u

Page 20 - Leitfähigkeitsmessgerät

4. Auswertung Der abgelesene Messwert stellt den Widerstand R [M] der Flüssigkeit zwischen den Elektroden (im Ringspalt) da

Page 21

5. Kalibrierung Das Gerät ist auf den Faktor 1.000 (d.h.1 M bzw. 1.00 S) eingestellt. Für die Kalibrierung ist ein Mess-widerstand von 1 M ± 1 %

Page 22

Der angezeigte Wert kann mit der M-Taste in Schritten von 0.005 verkleinert und mit der S-Taste in Schritten von 0.005 vergrößert werden. Wird eine

Page 24 - 2. Systembeschreibung

6. Störungen und Störungsbehebung Störung Beim Auslösen der Messung (Drücken der Messtaste) erscheint "Io.bA"

Page 25

Störung Nach Abschluss der 4 sec. Vorbereitungszeit bleibt die Anzeige weiterhin auf "Überlauf": Störungsbehebung • Anschluss des Messkabel

Page 26 - Eintauchtiefe

7. Lieferhinweise/ Technische Daten Best.Nr. PW-1710 Leitfähigkeits-Messzelle mit Verbindungskabel ASTM D 5682 Abmessung

Page 27 - 7,55 • 10

Best.Nr. PW-1722 Leitfähigkeitsmessgerät BYK-LC 2 mit Digitalanzeige ASTM D 5682 Messbereich: 0.05 -19.99 M 9.99 - 0.05 S Auflösung: 0.01 M

Page 28 - 5. Kalibrierung

198 013 256 EDI 0806 Conduttivimetro LC 2 N° di cat. PW-1722,

Page 30 - Störungsbehebung

Conduttivimetro LC 2 N° di cat PW-1722, PW-1710, PW-1712 Indice • Informazioni di sicurezza • Descrizione del sistema • Start-

Page 32 - Länge: ca. 250 mm

Conduttivimetro LC 2 N° di cat. PW-1710 Conduttivimetro LC 2 N° di cat. PW-1722 Conduttivimetro LC

Page 33

1. Informazioni di sicurezza • Lo strumento non deve essere usato in aree a prova di esplosione! • Usare solo ba

Page 34

Conductivity Meter LC 2 Cat.No. PW-1722, PW-1710, PW-1712 Table of Contents • Safety Information • System Description • Start-Up

Page 35

2. Descrizione del sistema Il Conduttivimetro LC 2, ASTM D 5682, è usato per l’analisi di liquidi a bassa conducibil

Page 36 - Conduttivimetro

3. Start up/ procedura di misurazione • Controllare che la cella di misurazione sia pulita; assemblare. Il solido design assi

Page 37

Nota Ogni volta che si effettua una misurazione, viene effettuato anche un auto-test. Durante la misurazione tutti i segmenti vengono mostrati sul di

Page 38

4. Valutazione Il valore misurato è la resistenza R [M] del liquido tra gli elettrodi (nello spazio tra gli anelli). La c

Page 39 - sicurezza

5. Calibrazione Lo strumento è presettato ad un fattore di 1.000 (cioè1 M o 1.00 S). Per la calibrazione servono un resistore di 1 M ± 1% o una

Page 40 - sistema

Il valore sul display può essere diminuito di 0.005 premendo il tasto M o aumentato di 0.005 con il tasto S. Se un tasto viene premuto a lungo, il v

Page 41

6. Ricerca dei problemi Sintomo All’inizio della misurazione (premendo il tasto per la misurazione) appare sul display “

Page 42 - Sonda piatta

Sintomo Dopo il termine dei 10 secondi del tempo di preparazione, il display continua a mostrare “overflow”: Possibile causa e rimedio • Controllare

Page 43

7. Guida per l’ordinazione / Specifiche tecniche N° di cat. PW-1710 Elettrodo a tubo per conducibilità con cavo di connessione

Page 44 - 5. Calibrazione

N° di cat. PW-1722 Conduttivimetro BYK LC 2 con display digitale ASTM D 5682 Range di misura: 0.05 - 19.99 M 9.99 - 0.05 S Risoluzione: 0.0

Page 46 - 6. Ricerca dei

EG - Konformitätserklärung EC - Declaration of Conformity Wir / We, BYK-Ga

Page 47

Déclaration de conformité - Nous, BYK-Gardner GmbH, déclarons que les produits / instruments ci-dessus mentionnés ont été développés, produits et cons

Page 48

BYK-Gardner GmbH Lausitzer Str. 8 D-82538 Geretsried Germany Tel. 0-800-gardner (0-800-4273637) +49-8171-3493-0 Fa

Page 50

198 013 256 EDI 0806

Page 51

Conductivity Tup Electrode Cat.No. PW-1710 Conductivity Meter BYK LC 2 Cat.No. PW-1722 Conductivity

Page 52

1. Safety Instructions • The instrument must not be used in explosion-proof areas! • Only use leak-proof batteries

Page 53

2. System Description The Conductivity Meter LC 2, ASTM D 5682, is used for the analysis of liquids of low conductivity,

Page 54 - 198 013 256 EDI 0806

3. Start-up/ Measuring Porcedure • Check measuring cell for clean condition; assemble. The solid design ensures the firm s