reference
set up
SET UP MENUE
Calibrate
Print Statistic
Print Online: off
Cont. Mode: on
Language: USA
Escape: SET UP
Calibration
Zero Calibration
Please cover
haze-port
Ready: OPERATE
Escape: SET UP
Calibration
100% Calibration
Please remove cover
from haze-port
Ready: OPERATE
Calibration
Place Transm.stand.
86.3
at haze-port
Ready: OPERATE
Escape: SET UP
Call up the
SET-UP menu
Call up the
calibration
routine
Cover the haze port with the
protective cover/dark standard, and
press the <operate> button to start
the measurement.
The bright measurement follows
after the dark measurement. The
bright measurement is referred to the
amount of light beamed in by
THEILLUMINATIONUNIT