Calibrate
7.2 Calibration Notes
The standards should be cleaned periodically.
For cleaning procedure see section "Cleaning
and Maintenance“. Please make sure that the
standards are not scratched.
Do not move the instrument while taking a
calibration measurement. If motion is detected,
an error message will be displayed and
calibration is aborted.
When moving from cold to warm environment,
there is a danger of condensation. For this
reason, you should wait for an appropriate
amount of time to allow the optical components
to adjust before calibrating and using the unit.